Specifications
Ventilation modes:
IPPV; A/C; SIMV;
Ventilator parameter ranges
Flow meter
O2(0.1-10L/min)
N2O(0.1-10L/min)
AIR(0.1-10L/min)
Rapid oxygen supply
35L/min-75L/min
Tidal volume(Vt)
0, 20 mL ~ 1500 mL
Frequence(Freq)
1bpm ~ 100 bpm
I/E
4:1~1:8
PEEP
0cmH2O ~ 30 cmH2O
Pressure triggering sensitivity(PTr)
-20 cmH2O ~ 0 cmH2O(Based on PEEP)
Flow trigger sensitivity (FTr)
0.5 L/min ~ 30 L/min
Pressure control (PC)
5 cmH2O ~ 60 cmH2O
SIGH
0(off) 1/100 ~ 5/100
Apnea Ventilation
OFF,5 s ~ 60 s
Pressure Limit
20 cmH2O ~ 100 cmH2O
Monitored Parameters
Frequence(Freq)
0 /min ~ 100 /min
Tidal volume(Vt)
0 mL ~ 2000 mL
MV
0 L/min ~ 100 L/min
Oxygen concentration
15 % ~ 100 %
Graphical display:
Oxygen concentration
Size
Wooden case packing size:L 560*W 820*H 1450mm , GW:104KG ; NW:64KG
Alarm and protection
No tidal volume
≤5Ml within 6s
High oxygen concentration alarm
Low oxygen concentration alarm
19%-100%
18%-99%
High Airway pressure alarm
Low Airway pressure alarm
High Minute Volume alarm
Low Minute Volume alarm
Continuous Pressure alarm
20cmH2O-100cmH2O
0cmH2O-20cmH2O
Adult(5L/min-20L/min) Paed(1L/min-15L/min
0-10L/min)
(PEEP+1.5kPa) over 16s
Suffocation warning
5s-60s no spontaneous ventilation
The maximum limited pressure
<12.5 kPa
Fan error
Oxygen deficit
Show on screen
Show on screen
Working conditions
Gas source
O2,N2O
Pressure
280kPa-600kPa
Voltage
100-240V
Power frequency
50/60Hz
Input power
40VA
Vaporizer
Reviews